StrainMatic®
Function
Automatic and Objective Measurement of the
Stress Birefringence in Glass and Plastics
and other Translucent Materials

Features
and
Benefits
- Automatic determination and color-coded presentation of the stress distribution
- Objective results, fast and easy operation
- Improvement in quality by on-site measurement
- Traceability by automatic filing of all parameters and results
- Cost reduction by optimization of the production process
Applications
- Optical materials and components (stress birefringence, homogeneity, micro-defects)
- Container glass (annealing lehr optimization, energy saving)
- Lighting (residual stress, glass metal transition)
- Tubing and pharmaceutical glass (residual stress)
- Plastic films (workability, polarization)
User Interface
with Result of a
Measurement

StrainMatic® M3
(left)
StrainMatic® M2
(right)
Technical
Data
| StrainMatic® M3/M | StrainMatic® M2/M | |
| main application | glass and plastic products | container glass |
| operation | external PC with TFT monitor, mouse, keyb. | integrated IPC with touch-screen |
| metering chamber | approx. 465 x 450 x 350 mm (H/W/D) | approx. 450 x 450 x 350 mm (H/W/D) |
| operating distance | approx. 465 mm (fixed) | approx. 120 - 450 mm, motor adjustment |
| illumination | LED array, approx. 240 x 200 mm | LED array, approx. 240 x 200 mm |
| image acquisition | matrix camera (640 x 480 pixels) | matrix camera (640 x 480 pixels) |
| image size | max. 200 x 150 mm (with 16 mm lens) | max. 200 x 150 mm (with 16 mm lens) |
| measuring area | variable rectangular, round or elliptical | variable rectangular, round or elliptical |
| measuring results |
polarization angle (°) temper number (ASTM C 148-00) optical retardation (nm) normalized optical retardation (nm/cm) integrated stress (MPa) |
polarization angle (°) temper number (ASTM C 148-00) optical retardation (nm) normalized optical retardation (nm/cm) integrated stress (MPa) |
| measuring range |
approx. -290 to +290 nm optical retardation optional -2900 to +2900 nm (HOD module) |
approx. -290 to +290 nm optical retardation optional -2900 to +2900 nm (HOD module) |
| repeatability | typical sigma(RMS) < 0.2 nm | typical sigma(RMS) < 0.2 nm |
| interfaces | Ethernet, USB, VGA | Ethernet, USB, VGA |
| power supply | 230 V, 50 Hz or 115 V, 60 Hz | 230 V, 50 Hz or 115 V, 60 Hz |
| dimensions | approx. 930 x 550 x 550 mm (H/W/D) | approx. 1200 x 550 x 550 mm (H/W/D) |
StrainMatic®
M4/130 (left)
StrainMatic®
M4/50 (right)


Technical
Data
| StrainMatic® M4/130 | StrainMatic® M4/50 | |
| main application | optical materials and components | optical materials, components, glass tubes |
| operation | external PC with TFT monitor, mouse, keyb. | external PC with TFT monitor, mouse, keyb. |
| metering chamber | approx. 240 x 450 x 450 mm (H/W/D) | approx. 150 x 350 x 350 mm (H/W/D) |
| illumination | LED array, approx. 200 x 160 mm | LED array, approx. 60 x 50 mm |
| image acquisition | matrix camera (640 x 480 pixels) with telecentric lens (130 mm aperture) | matrix camera (640 x 480 pixels) with telecentric lens (50 mm aperture) |
| image size | approx. 104 x 78 mm (0.16 mm/pixel) | approx. 40 x 30 mm (0.06 mm/pixel) |
| measuring area | variable rectangular, round or elliptical | variable rectangular, round or elliptical |
| measuring results |
polarization angle (°) optical retardation (nm) normalized optical retardation (nm/cm) integrated stress (MPa) |
polarization angle (°) optical retardation (nm) normalized optical retardation (nm/cm) integrated stress (MPa) |
| measuring range |
approx. -290 to +290 nm optical retardation optional -2900 to +2900 nm (HOD module) |
approx. -290 to +290 nm optical retardation |
| repeatability | typical sigma(RMS) < 0.1 nm | typical sigma(RMS) < 0.1 nm |
| interfaces | Ethernet, USB, VGA | Ethernet, USB, VGA |
| power supply | 230 V, 50 Hz or 115 V, 60 Hz | 230 V, 50 Hz or 115 V, 60 Hz |
| dimensions | approx. 935 x 550 x 550 mm (H/W/D) | approx. 710 x 440 x 440 mm (H/W/D) |
StrainMatic®
M4/60 zoom (left)
StrainMatic®
M4/50 tube tester
(right)


Technical
Data
| StrainMatic® M4/60 zoom | StrainMatic® M4/50 tube tester | ||
| main application | optical materials and components | tube glass | |
| operation | internal PC with TFT monitor, mouse, keyb. | internal PC with touch-screen | |
| metering chamber | approx. 150 x 550 x 550 mm (H/W/D) | cell approx. 100 x 50 x 50 mm (H/B/T) | |
| illumination | LED array, approx. 120 x 100 mm | LED array, approx. 50 x 60 mm | |
| image acquisition | matrix camera (1280 x 1024 pixels) with telecentric 4x zoom objective lens | matrix camera (1280 x 1024 pixels) with telecentric lens (50 mm aperture) | |
| image size |
approx. 48 x 36 mm (0.04 mm/pixel) to approx. 12 x 9 mm (0.01 mm/pixel) |
approx. 30 x 38 mm (0.03 mm/pixel) | |
| measuring area | variable rectangular, round or elliptical | automatic or manually adjustable | |
| measuring results |
polarization angle (°) optical retardation (nm) normalized optical retardation (nm/cm) integrated stress (MPa) |
polarization angle (°) optical retardation (nm) normalized optical retardation (nm/cm) tube wall stress (MPa) |
|
| measuring range | approx. -290 to +290 nm optical retardation | approx. -290 to +290 nm optical retardation | |
| repeatability | typical sigma(RMS) < 0.2 nm | typical sigma(RMS) < 0.2 nm | |
| interfaces | Ethernet, USB, DVI | Ethernet, USB, DVI | |
| power supply | 230 V, 50 Hz or 115 V, 60 Hz | 230 V, 50 Hz or 115 V, 60 Hz | |
| dimensions | approx. 1150 x 650 x 650 mm (H/W/D) | approx. 1360 x 800 x 480 mm (H/W/D) |


StrainMatic®
M4/160 Stepper (left)
StrainMatic®
M4/300 (right)
Technical
Data
| StrainMatic® M4/160.42 Stepper | StrainMatic® M4/300 | |
| main application | optical materials and components | optical materials and components |
| operation | integrated PC with touch-screen | external PC with TFT monitor, mouse, keyb. |
| sample positioning | automatic XY table | manual XY table with telescopic slide-out |
| illumination | LED, collimated or 5° divergent | LED array, approx. 400 x 320 mm |
| image acquisition | matrix camera (2048 x 2048 pixels) with telecentric lens (160 mm aperture) | matrix camera (640 x 480 pixels) with telecentric lens (300 mm aperture) |
| image size | approx. 112 x 112 mm (0.06 mm/pixel) | approx. 232 x 174 mm (0.36 mm/pixel) |
| measuring area | variable rectangular, round or elliptical | variable rectangular, round or elliptical |
| measuring results |
polarization angle (°) optical retardation (nm) normalized retardation (nm/cm) integrated stress (MPa) |
polarization angle (°) optical retardation (nm) normalized retardation (nm/cm) integrated stress (MPa) |
| measuring range |
approx. -68 to +68 nm (standard)
approx. -270 to +270 nm (extended) |
approx. -290 to +290 nm optical retardation
optional -2900 to +2900 nm (HOD module) |
| repeatability | typical sigma(RMS) < 0.02 nm | typical sigma(RMS) < 0.1 nm |
| interfaces | Ethernet, USB, DVI | Ethernet, USB, VGA |
| power supply | 230 V, 50 Hz or 115 V, 60 Hz | 230 V, 50 Hz or 115 V, 60 Hz |
| dimensions | approx. 1850 x 1800 x 1500 mm (H/W/D) | approx. 2210 x 1050 x 850 mm (H/W/D) |
References
Application
Examples


























, Internet: www.ilis.de